4Hacettepe Universty School of Medicine, Department of Radiology, Ankara, Turkey DOI : 10.5137/1019-5149.JTN.25776-19.2 AIM: To investigate the safety and efficacy of Derivo® embolisation device (DED), a new-generation flow diverter designed to treat cerebrovascular aneurysms, and its long-term clinical outcomes.
MATERIAL and METHODS: In total, 146 patients with 182 aneurysms were treated with DED. The mean age of the participants was 51.5 years; among them, 46 (31.5%) presented with acute subarachnoid haemorrhage. The mean aneurysm size was 8.3 mm, and 12 aneurysms were involved the vertebrobasilar system. Ophthalmic aneurysms account for most internal carotid artery (ICA) aneurysms.
RESULTS: The Glasgow Coma Scale (GCS) score of 12 patients was <15. DED was associated with a mortality rate of 2.7% and permanent morbidity rate of 3.4%, and a complete aneurysm occlusion rate was achieved in 78.7% of cases after 7.02 months.
CONCLUSION: The DED device is a new-generation flow diverter with excellent opening behaviour and navigational benefits. Our results indicated a safe aneurysm occlusion with optimum morbidity and mortality values despite the fact that almost one-third of the patients presented with subarachnoid haemorrhage.